| Displays & Controls: |
| Main Display |
10.4" high-resolution active matrix color LCD
Resolution = 640 x 480 pixels |
| Function Keys |
ON/OFF, BP Zero All, NIBP start/stop, PRINT start/stop, Alarm Silence, rotary encoded push-button control knob |
| Indicators: |
| AC Power |
Green LED w/power plug icon |
| Battery Power |
Green LED w/battery icon |
| SpO2: |
| Range |
0-100% (function saturation) |
| Accuracy |
±2% @ 70-100% ±3% @ 50-69% |
| Alarm Range |
50-100% and OFF |
| Averaging |
4, 8, 16 pulse beat average |
| Pulse Rate (SpO2) or Heart Rate (ECG): |
| Range |
30-250 bpm (SpO2) |
| 30-180 bpm (NIBP) |
| 30-300 bpm (ECG) |
| Accuracy |
±2% bpm or ±2% (whichever is greater) |
| Alarm Range |
50-100% and OFF |
| Averaging |
Fixed at 8 seconds |
| ECG: |
| Configuration |
3-lead (standard) or 5 lead (optional) |
| Lead Selection |
I, II, III, aVL, aVR, aVF and V |
| Display Gain |
X1/2, X1, X2, X4 |
| Defib. Protection |
Yes |
| Pace Detect/Reject |
Yes |
| Sweep Speed |
25 mm/sec |
| Impedance Respiration (not available in neonate mode): |
| Measuring Method |
Impedance Pneumography |
| Rate |
3-150 breaths/min |
| Accuracy |
±1 breath/min or ±5% (whichever is greater) |
| Alarm Range |
5-150 bpm & OFF (1 bpm steps) |
| Non-Invasive Blood Pressure: |
| Range |
20-250 mmHg (adult/ped) |
| 10-135 mmHg (neonate) |
| Accuracy |
Meets or exceeds SP10-1992 Standard for NIBP accuracy |
| Cuff Pressure Accuracy |
±3 mmHg or ±2% (whichever is greater) |
| Inflation Pressure Adjustment Range |
50-300 mmHg (adult/ped) |
| 50-150 mmHg (neonate) |
| Default Inflation Pressure |
175 mmHg (adult) 100 mmHg (neonate) |
| Alarm Range |
20-300mmHg & OFF (1mmHg steps) (adult/ped) |
| 20-135mmHg & OFF (1mmHg steps) (neonate) |
| 2 Channel Invasive Pressure - optional: |
| Range |
-50 mmHg - 300 mmHg (after zeroing) |
| Accuracy |
±1 mmHg or ±1% (whichever is greater) |
| Alarm Range |
-50 - 300 mmHg (in 1 mmHg steps) |
| 2 Channel Temperature - optional: |
| Range |
0-50°C (32-122°F) |
| Accuracy |
±0.1°C (±0.2°F) plus temp sensor tolerance |
| Resolution |
0.1°C (0.1°F) |
| Alarm Range |
20.0-50.0°C (68-122°F) & OFF |
| Sensor Compatibility |
YSI 400 or 700 series |
| Sidestream Capnography - optional (not available in neonate mode): |
| Measurement |
Non-Dispersive IR absorption |
| Measurement Range |
0-10% CO2 STPD |
| Display Range |
0-100 mmHg; 0-13.3 kPa; 0-10% CO2 |
| Accuracy |
±2mmHg or 4% of reading (whichever is greater) (0-10% CO2) |
| Alarm Range |
0-100 mmHg; 0-13.3 kPa; 0-10% CO2 & OFF |
| N2O Compensation |
selectable 40% (default = OFF) |
| Averaging |
4 breath |
| Flow Rate |
140 ml/min± 20 ml |
| Printer - optional: |
| Dual trace printer provides waveforms, annotation lines, snapshot, continuous, print on alarms and trend tables |
| Dimensions: |
| 12" H x 13-1/4" W x 7" D (30.5cm H x 34.0cm W x 18.0cm D) |
| Weight: |
| Without battery or capnograph |
12 pounds (5.4 kg) |
| With battery & capnograph |
15 pounds (6.8 kg) |
| Power Requirements: |
| AC Input |
90 - 120 VAC, 230-240 VAC, 50/60 Hz, 0.6 A |
| Battery (optional) |
1 hour operation |
| Auxiliary Inputs/Outputs: |
| Serial RS232 standard |
| Optional analog output to interface with the Life Sensing Instrument Company, Inc. (LSI) Central Station |
| Environmental Specifications: |
| Operating Temperature |
0-40°C |
| Storage Temperature |
-40°-75°C (-40°-167°F) |
| Relative Humidity |
15-95% (operating) |
| (non-condensing) |
10-95% (storage) |
| Standards: |
| EN60601-1, EN60601-1-1, UL 2601, CSA Std C22.2 No. 601.1, EN60601-1-2 |
| Regulatory Compliance: |
510(k): 982279, K010770, K011177, K030407
ISO 9001 / EN 46001 |
|
| 3-LEAD OR 5-LEAD SELECTABLE ECG |
| Heart Rate Range: |
20-250 bpm |
| Heart Rate Accuracy: |
± 3 bpm or ± 3%, whichever is greater |
| Bandwidth: |
0.5-40 Hz (monitor mode) or 0.05-100 Hz (extended mode) |
| Display Sweep Speeds: |
25, 12.5 or 6.25 mm/sec |
| NONINVASIVE BLOOD PRESSURE |
| Systolic Measurement Range: |
60-250 mmHg |
| Diastolic Measurement Range: |
30-160 mmHg |
| Blood Pressure Accuracy: |
Meets AAMI SP10-1992 Standard |
| Operating Modes: |
Manual (single determination) or Automatic Mode |
| Cycle Interval Choices: |
1, 3, 5, 10, 15, 30 or 60 minutes |
| PULSE OXIMETRY (NELLCOR PURITAN BENNETT™) |
| Measurement Range: |
40-100% |
| Accuracy: |
± 2 digits (no motion), ± 3 digits (with motion) in the range of 70-100% |
| Pulse Rate Range: |
20-250 bpm |
| Pulse Rate Accuracy: |
± 3 digits (no motion) at 20-250 bpm, ± 5 digits (with motion) at 55-125 bpm |
| TEMPERATURE |
| Range: |
17-50º C (62.6 -122.0º F) |
| Accuracy: |
Meets ASTM Accuracy Standards |
| Probe Type: |
YSI Series 400 Compatible |
| IMPEDANCE RESPIRATION |
| Range: |
5-100 breaths per minute |
| Accuracy: |
± 3 breaths per minute |
| END TIDAL CARBON DIOXIDE |
| Measurement Technique: |
Sidestream |
| Measurement Range: |
0-98 mmHg |
| ETCO2 Accuracy: |
0-40 mmHg: ± 3 mmHg; 41-98 mmHg ± 10% of reading |
| Breath Rate Measurement Range: |
5-100 breaths per minute |
| Breath Rate Accuracy: |
± 1 breath per minute or ± 5%, whichever is greater |
| GENERAL |
| Size: |
35 cm x 23 cm x 25 cm (13.8" x 9.1" x 9.8") |
| Weight: |
4.3-5.5 kg (9.5 -12.1 lbs) dependant on configuration |
| Power Requirements: |
120 VAC, 50/60 Hz; 240 VAC, 50/60 Hz |
| Printer: |
Thermal array, prints waveforms and numerical trends |
| Battery: |
Lead acid, rechargeable |
| ORDERING INFORMATION |
| MODEL |
DESCRIPTION |
| FGI-621NO |
ECG / Nellcor Puritan Bennett SpO2 / NIBP |
| FGI-621NP |
ECG / Nellcor Puritan Bennett SpO2 / NIBP / Printer |
| FGI-622N0 |
ECG / Nellcor Puritan Bennett SpO2 / NIBP / Temperature / Respiration / Battery / RS232 |
| FGI-622NP |
ECG / Nellcor Puritan Bennett SpO2 / NIBP / Temperature / Respiration / Battery / RS232 / Printer |
| FGI-623NP |
ECG / Nellcor Puritan Bennett SpO2 / NIBP / ETCO2 / Temperature / Respiration / Battery / RS232 / Printer |
|
| Display: |
| 16 character liquid crystal display |
| Dimensions: |
| Height: 6.6 in., Width: 10.0 in., Depth: 6.0 in. |
| 16.8 cm x 25.4 cm x 15.2 cm |
| Weight: |
5.4 lbs (2.4 kg) |
| Blood Pressure Accuracy: |
| Meets or exceeds SP10-2002 AAMI standards |
| Blood Pressure Determination Time: |
| 20-45 seconds, typical |
| Cuff Pressure Range: |
| 0 to 270 mmHg, 0 to 36 kPa |
| Initial Cuff Inflation: |
| Adult: |
160 mmHg (21.3 kPa) |
| Pediatric: |
120 mmHg (16 kPa) |
| Neonate: |
90mmHg (12kPa) |
| Systolic Range: |
| Adult: |
30 to 260 mmHg (4 to 34.5 kPa) |
| Pediatric: |
30 to 160 mmHg (4 to 21.3 kPa) |
| Neonate: |
25 to 120 mmHg (3.3 to 16 kPa) |
| Diastolic Range: |
| Adult: |
20 to 235 mmHg (2.7 to 31.3 kPa) |
| Pediatric: |
15 to 130 mmHg (2 to 17.3 kPa) |
| Neonate: |
10 to 105 mmHg (1.3 to 14 kPa) |
| Overpressure Cutoff: |
| 295 to 330 mmHg (39.3 to 44 kPa) |
| Pulse Rate Range: |
| 30 to 220 bpm |
| Pulse Rate Accuracy: |
| ± 3 bpm or 3% |
| SpO2 |
| Saturation Range: |
| 1% to 100% |
| Accuracy: |
| Adult/Pediatric: |
70-100% ±2 digits |
| Neonate: |
70-100% ±3 digits |
| Pulse Rate Range: |
| 20 to 255 bpm |
| Pulse Rate Accuracy: |
| ±2 bpm (without motion) |
| ±3 bpm (with motion) |
| ±3 digits (low perfusion) |
| Temperature: |
| Measurement Range: |
| 80F to 110F (26.7C to 43.3C) |
| Temperature Calibration Accuracy: |
| ± 0.2F (±0.1C); meets or exceeds ASTME1112-00; EN12470-3:2000 |
| Determination Time (approximate): |
| Predictive |
| Oral: |
4-6 seconds |
| Pediatric Axillary: |
10-13 seconds |
| Adult Axillary: |
10-13 seconds |
| Rectal: |
10-13 seconds |
| Monitored |
| Oral/Rectal: |
3 minutes |
| Axillary: |
5 minutes |
| Printer: |
| Thermal transfer printer |
| Power: |
| AC Power supply: 8 volts DC and 0.75 amperes |
| Rechargeable sealed lead acid internal battery |
| Operating time: 8 hours (Adult NIBP monitoring, at 3-minute intervals, with simultaneous and continuous monitoring of temperature and SpO2 values |
| Battery recharge time: 90% in 12 hours |
| Simultaneously operates monitor and charges battery when AC power adapter is plugged in |
| ORDERING INFORMATION |
| MODEL |
DESCRIPTION |
| FGI-53000 |
NIBP only (includes Pulse Rate and MAP) |
| FGI-5300P |
NIBP/ Printer |
| FGI-530T0 |
NIBP/ Temperature |
| FGI-530TP |
NIBP/ Temperature / Printer |
| FGI-53N00 |
NIBP/ Pulse Oximetry |
| FGI-53N0P |
NIBP/ Pulse Oximetry/ Printer |
| FGI-53NT0 |
NIBP/ Pulse Oximetry / Temperature |
| FGI-53NTP |
NIBP/ Pulse Oximetry / Temperature / Printer |
| 008-0835-00 |
Mobile Stand with Accessory Basket |
| 008-0834-00 |
Wall Mount with Accessory Basket |
|
| Dimensions: |
| 6.5 in (16.5 cm) x 8.6 in (21.8 cm) x 5.0 in (12.7 cm) |
| Weight: |
6 lb (2.8 kg) |
| Blood Pressure Accuracy: |
| Meets or exceeds SP10-1992 AAMI standards |
| Cuff Pressure Range: |
| 0-300 mmHg |
| Systolic Range: |
| Max: |
250 mmHg |
| Min: |
60 mmHg |
| Diastolic Range: |
| Max: |
160 mmHg |
| Min: |
30 mmHg |
| Pulse Rate Accuracy (using Blood Pressure determination): ±5% |
| Pulse Rate Accuracy (using SpO2 determination): |
| ±3 digits (without motion) |
| 55 - 125bpm, ±5 digits (with motion) |
| ±3 digits (low perfusion) |
| Pulse Rate Range: |
| Max: |
200 bpm |
| Min: |
40 bpm |
| Temperature Calibration Accuracy: |
| ±0.2°F (±0.1°C) |
| Temperature Range: |
| Normal and Monitor Mode |
| Max: |
108.0°F (42.2°C) |
| Min: |
84.0°F (28.9°C) |
| Pulse Oximeter Accuracy: |
| Without Motion - Adults: |
70 to 99%, ±2 digits |
| With Motion - Adults: |
70 to 99%, ±3 digits |
| Low Perfusion: |
70 to 99%, ±2 digits, <70% unspecified |
| Printer: |
| Thermal transfer printer |
| Power Requirements: |
| North American Version: |
120 VAC, 60 Hz |
| International Version: |
220-240 VAC, 50-60 Hz |
| Internal Battery: |
Lead acid, with external recharger |
| ORDERING INFORMATION |
| MODEL |
DESCRIPTION |
| FGI-52000 |
NIBP only (includes Pulse Rate and MAP) |
| FGI-5200P |
NIBP/ Printer |
| FGI-520T0 |
NIBP/ Temperature |
| FGI-520TP |
NIBP/ Temperature / Printer |
| FGI-52N00 |
NIBP/ Pulse Oximetry |
| FGI-52N0P |
NIBP/ Pulse Oximetry/ Printer |
| FGI-52NT0 |
NIBP/ Pulse Oximetry / Temperature |
| FGI-52NTP |
NIBP/ Pulse Oximetry / Temperature / Printer |
|
| Dimensions: |
| h: 9.70”, l: 5.72”, w: 4.73” |
| h: 24.64 cm, l: 14.53 cm, w: 12.01 cm |
| Weight: |
4.5 lbs. (2.04 kg) |
| Blood Pressure Accuracy: |
Meets or exceeds SP10-1992 AAMI standards |
| Cuff Pressure Range: |
0-300 mmHg |
| Systolic Determination: |
| Max: |
250 mmHg |
| Min: |
60 mmHg |
| Diastolic Determination: |
| Max: |
160 mmHg |
| Min: |
30mmHg |
| Pulse Rate Accuracy: |
| SpO2 Module Heart Rate: |
± 3.0 bpm |
| Blood Pressure Algorithm: |
± 5.0 bpm |
| Pulse Rate Determination: |
Max: 245 bpm Min: 25 bpm |
| Temperature Accuracy: |
Meets or exceeds ASTM E112-86 standards |
| Pulse Oximeter Accuracy |
| Nellcor Puritan Bennett Pulse Oximeter Module: |
± 3% in the range of 70-100% Oxygen saturation (1 standard deviation) |
| < 70% unspecified by the OEM |
| Power Requirements: |
| North American Version: |
120 VAC, 60 Hz |
| International Version: |
230-240 VAC, 50 Hz |
| Internal Battery: |
Lead Acid, with external charger |
| ORDERING INFORMATION |
| MODEL |
DESCRIPTION |
| FGI-4200B-E1 |
NIBP only (pulse rate and MAP) |
| FGI-420TB-E1 |
NIBP/Temperature |
| FGI-42N0B-E1 |
NIBP/Nellcor Pulse Oximetry |
| FGI-42NTB-E1 |
NIBP/Nellcor Pulse Oximetry/Temperature |
|