| 3-LEAD OR 5-LEAD SELECTABLE ECG |
| Heart Rate Range: |
20-250 bpm |
| Heart Rate Accuracy: |
± 3 bpm or ± 3%, whichever is greater |
| Bandwidth: |
0.5-40 Hz (monitor mode) or 0.05-100 Hz (extended mode) |
| Display Sweep Speeds: |
25, 12.5 or 6.25 mm/sec |
| NONINVASIVE BLOOD PRESSURE |
| Systolic Measurement Range: |
60-250 mmHg |
| Diastolic Measurement Range: |
30-160 mmHg |
| Blood Pressure Accuracy: |
Meets AAMI SP10-1992 Standard |
| Operating Modes: |
Manual (single determination) or Automatic Mode |
| Cycle Interval Choices: |
1, 3, 5, 10, 15, 30 or 60 minutes |
| PULSE OXIMETRY (NELLCOR PURITAN BENNETT™) |
| Measurement Range: |
40-100% |
| Accuracy: |
± 2 digits (no motion), ± 3 digits (with motion) in the range of 70-100% |
| Pulse Rate Range: |
20-250 bpm |
| Pulse Rate Accuracy: |
± 3 digits (no motion) at 20-250 bpm, ± 5 digits (with motion) at 55-125 bpm |
| TEMPERATURE |
| Range: |
17-50º C (62.6 -122.0º F) |
| Accuracy: |
Meets ASTM Accuracy Standards |
| Probe Type: |
YSI Series 400 Compatible |
| IMPEDANCE RESPIRATION |
| Range: |
5-100 breaths per minute |
| Accuracy: |
± 3 breaths per minute |
| END TIDAL CARBON DIOXIDE |
| Measurement Technique: |
Sidestream |
| Measurement Range: |
0-98 mmHg |
| ETCO2 Accuracy: |
0-40 mmHg: ± 3 mmHg; 41-98 mmHg ± 10% of reading |
| Breath Rate Measurement Range: |
5-100 breaths per minute |
| Breath Rate Accuracy: |
± 1 breath per minute or ± 5%, whichever is greater |
| GENERAL |
| Size: |
35 cm x 23 cm x 25 cm (13.8" x 9.1" x 9.8") |
| Weight: |
4.3-5.5 kg (9.5 -12.1 lbs) dependant on configuration |
| Power Requirements: |
120 VAC, 50/60 Hz; 240 VAC, 50/60 Hz |
| Printer: |
Thermal array, prints waveforms and numerical trends |
| Battery: |
Lead acid, rechargeable |
| ORDERING INFORMATION |
| MODEL |
DESCRIPTION |
| FGI-621NO |
ECG / Nellcor Puritan Bennett SpO2 / NIBP |
| FGI-621NP |
ECG / Nellcor Puritan Bennett SpO2 / NIBP / Printer |
| FGI-622N0 |
ECG / Nellcor Puritan Bennett SpO2 / NIBP / Temperature / Respiration / Battery / RS232 |
| FGI-622NP |
ECG / Nellcor Puritan Bennett SpO2 / NIBP / Temperature / Respiration / Battery / RS232 / Printer |
| FGI-623NP |
ECG / Nellcor Puritan Bennett SpO2 / NIBP / ETCO2 / Temperature / Respiration / Battery / RS232 / Printer |