| Display: |
| 16 character liquid crystal display |
| Dimensions: |
| Height: 6.6 in., Width: 10.0 in., Depth: 6.0 in. |
| 16.8 cm x 25.4 cm x 15.2 cm |
| Weight: |
5.4 lbs (2.4 kg) |
| Blood Pressure Accuracy: |
| Meets or exceeds SP10-2002 AAMI standards |
| Blood Pressure Determination Time: |
| 20-45 seconds, typical |
| Cuff Pressure Range: |
| 0 to 270 mmHg, 0 to 36 kPa |
| Initial Cuff Inflation: |
| Adult: |
160 mmHg (21.3 kPa) |
| Pediatric: |
120 mmHg (16 kPa) |
| Neonate: |
90mmHg (12kPa) |
| Systolic Range: |
| Adult: |
30 to 260 mmHg (4 to 34.5 kPa) |
| Pediatric: |
30 to 160 mmHg (4 to 21.3 kPa) |
| Neonate: |
25 to 120 mmHg (3.3 to 16 kPa) |
| Diastolic Range: |
| Adult: |
20 to 235 mmHg (2.7 to 31.3 kPa) |
| Pediatric: |
15 to 130 mmHg (2 to 17.3 kPa) |
| Neonate: |
10 to 105 mmHg (1.3 to 14 kPa) |
| Overpressure Cutoff: |
| 295 to 330 mmHg (39.3 to 44 kPa) |
| Pulse Rate Range: |
| 30 to 220 bpm |
| Pulse Rate Accuracy: |
| ± 3 bpm or 3% |
| SpO2 |
| Saturation Range: |
| 1% to 100% |
| Accuracy: |
| Adult/Pediatric: |
70-100% ±2 digits |
| Neonate: |
70-100% ±3 digits |
| Pulse Rate Range: |
| 20 to 255 bpm |
| Pulse Rate Accuracy: |
| ±2 bpm (without motion) |
| ±3 bpm (with motion) |
| ±3 digits (low perfusion) |
| Temperature: |
| Measurement Range: |
| 80F to 110F (26.7C to 43.3C) |
| Temperature Calibration Accuracy: |
| ± 0.2F (±0.1C); meets or exceeds ASTME1112-00; EN12470-3:2000 |
| Determination Time (approximate): |
| Predictive |
| Oral: |
4-6 seconds |
| Pediatric Axillary: |
10-13 seconds |
| Adult Axillary: |
10-13 seconds |
| Rectal: |
10-13 seconds |
| Monitored |
| Oral/Rectal: |
3 minutes |
| Axillary: |
5 minutes |
| Printer: |
| Thermal transfer printer |
| Power: |
| AC Power supply: 8 volts DC and 0.75 amperes |
| Rechargeable sealed lead acid internal battery |
| Operating time: 8 hours (Adult NIBP monitoring, at 3-minute intervals, with simultaneous and continuous monitoring of temperature and SpO2 values |
| Battery recharge time: 90% in 12 hours |
| Simultaneously operates monitor and charges battery when AC power adapter is plugged in |
| ORDERING INFORMATION |
| MODEL |
DESCRIPTION |
| FGI-53000 |
NIBP only (includes Pulse Rate and MAP) |
| FGI-5300P |
NIBP/ Printer |
| FGI-530T0 |
NIBP/ Temperature |
| FGI-530TP |
NIBP/ Temperature / Printer |
| FGI-53N00 |
NIBP/ Pulse Oximetry |
| FGI-53N0P |
NIBP/ Pulse Oximetry/ Printer |
| FGI-53NT0 |
NIBP/ Pulse Oximetry / Temperature |
| FGI-53NTP |
NIBP/ Pulse Oximetry / Temperature / Printer |
| 008-0835-00 |
Mobile Stand with Accessory Basket |
| 008-0834-00 |
Wall Mount with Accessory Basket |