System Spec
|
Application Area : Gallstone and urinary
track stones
Ston Localization identification : C-ARM X-Ray & Ultrasound
System |
|
Shock Wave
Generator (Model : SWG-5000)
|
Shock Wave Principle : Ellipsoid with spark gap discharge
Distance from Reflector Surface : 135mm
Patient Coupling : Flexible rubber membrane
Focal Pressure at target : 400-1100 bar
Voltage Range : 10-22KV (Adjustablel)
Shock Wave Impulse Delay Time : 1.0-9.9 SEC (Usually
1.0-2.0 SEC)
Trigger Mode :Auto or manual
Spark Gap :Permanent special type E-250L
|
|
Treatment Table
|
Table Top : Three peice flat matteress 2100
x 600
(1200 x600 when folded) |
|
(Model : KTT-5000)
|
Longitudinal Adjustment : 150mm
Transverse Adjustment : 150mm
Heigh Adjustment : 300 mm
Table move speed adjustment is possible |
|
C-Arm X-ray System
(Modes :KMC-5000A)
|
Fluoroscopy from 40-120KVP from 05.-4.0mA
(Boost mode Output : 20mA fIxed)
Tube Type :anode rotating
Focal size & Heating Unit : 0.3/0.6mm & 300,000HU
Image Intensifier : 6' or 9' (option)
C-arm rotation : ±30 degrees
Image Memory & Noise Reduction
SID (Tube to Intensifire) : 950mm |
|
Control Desk
|
Dimension : 720(W) x780(D)X1280(H)
Software Control : Shock Wave, X-ray exposure, patients, Record
setting
PC system : CPU-686, 4.0 G Hard,CD-ROM, 3.5" FDD,
15"
Monitor, Others
System contrl type : Programmable Logic Contrl (FARA-N70)
|
|
Water control
|
Water Capacity :2-3 Liter
Water control : Water in/Out, Air In/Out |
|
Ultrasound System
(Model : SA-5000 etc.)
|
Image Mode :B Mode, B/B Mode, B/M Mode, M Mode
Applications : Abdominal, Ob/Gyn, Cardiac, Vascular,
etc
Transducers : Linear Array, Convex Array
System Architecture : 12' Dual Port Connetors for 2
Transducers
Physical Demension : 450 (D)x668(W)x1165(H/Min)
|
|
General
|
Input Power Supply : 200-240 VAC /100-120VAC,
50-60 Hz
Power Consumption :3,000 VA
Electrical Safety : IEC 601.1, Glass I, TypeB
Overall Weight : 320 kg
Control Desk Weight :1 50 kg |
These specifications may be changed without notice.