First Gulf International, Inc.

First Gulf International, Inc.
Site Map
Complete Line of Medical Equipment & Hospital Package Deals
spacer1
spacer2

High Performance, Deep Impact with Compact Design E.S.W.L.
(Extracorporeal Shock Wave Lithotripter)


FOR PRICING AND DELIVERY, CALL US AT 713-961-7793 OR EMAIL US

FGI-SDS-5000 PLUS

SDS-5000 PLUS Shock Wave Generator
Shock Wave Generator
SDS-5000 PLUS Patient Table
Patient Table
SDS-5000 PLUS Operating Unit
Operating Unit
System Spec
Application Area : Gallstone and urinary track stones
Ston Localization identification : C-ARM X-Ray & Ultrasound System
Shock Wave
Generator
(Model : SWG-5000)

Shock Wave Principle : Ellipsoid with spark gap discharge
Distance from Reflector Surface : 135mm
Patient Coupling : Flexible rubber membrane
Focal Pressure at target : 400-1100 bar
Voltage Range : 10-22KV (Adjustablel)
Shock Wave Impulse Delay Time : 1.0-9.9 SEC (Usually 1.0-2.0 SEC)
Trigger Mode :Auto or manual
Spark Gap :Permanent special type E-250L

Treatment Table
Table Top : Three peice flat matteress 2100 x 600
(1200 x600 when folded)
(Model : KTT-5000)
Longitudinal Adjustment : 150mm
Transverse Adjustment : 150mm
Heigh Adjustment : 300 mm
Table move speed adjustment is possible
C-Arm X-ray System
(Modes :KMC-5000A)
Fluoroscopy from 40-120KVP from 05.-4.0mA
(Boost mode Output : 20mA fIxed)
Tube Type :anode rotating
Focal size & Heating Unit : 0.3/0.6mm & 300,000HU
Image Intensifier : 6' or 9' (option)
C-arm rotation : ±30 degrees
Image Memory & Noise Reduction
SID (Tube to Intensifire) : 950mm
Control Desk

Dimension : 720(W) x780(D)X1280(H)
Software Control : Shock Wave, X-ray exposure, patients, Record setting
PC system : CPU-686, 4.0 G Hard,CD-ROM, 3.5" FDD, 15"
Monitor, Others
System contrl type : Programmable Logic Contrl (FARA-N70)

Water control
Water Capacity :2-3 Liter
Water control : Water in/Out, Air In/Out
Ultrasound System
(Model : SA-5000 etc.)

Image Mode :B Mode, B/B Mode, B/M Mode, M Mode
Applications : Abdominal, Ob/Gyn, Cardiac, Vascular, etc
Transducers : Linear Array, Convex Array
System Architecture : 12' Dual Port Connetors for 2 Transducers
Physical Demension : 450 (D)x668(W)x1165(H/Min)

General
Input Power Supply : 200-240 VAC /100-120VAC, 50-60 Hz
Power Consumption :3,000 VA
Electrical Safety : IEC 601.1, Glass I, TypeB
Overall Weight : 320 kg
Control Desk Weight :1 50 kg
These specifications may be changed without notice.
©2007 First Gulf International, Inc. All Rights Reserved.

Powered by WebWorks USA - Website Hosting, Design and Promotion in Houston, Texas