| FGI Oximeter Specifications (Nominal) |
| Capnocheck® Sleep Capnograph/Oximeter |
|
| FGI-9004-051 |
| Displays, Indicators & Keys: |
| Display |
Vacuum Fluorescent |
| Waveform |
Selectable CO2 or CO2/Plethysmogram waveform |
| Graphic Trends |
Graphical display, factory selectable sample rate, user selectable trend up to 16 hr. display @ 12 sec. sample rate up to 10 hr. display @ 4 sec. sample rate |
| Pulse Strength |
Logarithmically scaled, 8-segment bar graph |
| Low Battery |
LED lights when approximately 30 minutes of battery life remain |
| Alarm Silence |
Alarm Select; Alarm Silence; UP arrow Alarm Limit/Pulse Volume Adjustment; DOWN arrow Alarm Limit/Pulse Volume Adjustment; |
| Sensor |
LED lights to indicate an Oximeter sensor problem |
| Lost Pulse Alarm |
Adjustable volume user-selectable high and low limits for all displayed parameters
Screen indicators for each parameter |
| SpO2: |
| Range |
0-100% |
| Accuracy |
±2% @ 70-100% |
| <70% unspecified |
| Averaging |
2, 4, 8, or 16-beat averaging (default: 2-beat) |
| Alarm Range |
High: 50-99% & OFF (1% steps) |
| Low: 50-98% & OFF (1% steps) |
| Pulse Rate: |
| Range |
30-254 bpm |
| Accuracy |
±2 bpm or ±2%, whichever is greater |
| Averaging |
8 second average |
| Alarm Range |
Pulse Tone - Pitch corresponds SpO2 value
Volume adjustable or OFF |
| Power Requirements: |
115V~, 60 Hz |
| (Optional) 100V~, 50 Hz |
| (Optional) 230V~, 50/60 Hz |
| Battery Life: |
Nickel Cadmium battery, 6 VDC, internal rechargeable, not user replaceable, fully charged continuous- use life 2 hours approximate, maximum full charge time 4 hours |
| Dimensions: |
3.5" H x 10.0" W x 5.5" D |
| 88.9 mm x 254 mm x 139.7 mm |
| Weight: |
4.8 pounds (2.2 kg) |
| Environment:; |
| Operating Temperature |
32°F - 122°F (0°C - 50°C) |
| Storage Temperature |
-40°F-158°F (-40°C-70°C) |
| Relative Humidity |
10 - 95% (storage) |
| (non condensing) |
15 - 95% (operating) |
| Auxiliary Serial Printer/Computer Output: |
Analog Output: Any four of: CO2 waveform, Plethysmogram, End Tidal CO2, Resiration Rate, Inspired CO2, % SpO2, Pulse Rate, 0V or 1V Cal Signal
Range: 0-1 VDC full scale |
Serial Output: RS232C: tabular trends, patient data
For Sleep Study: 10 hrs. @ 4 sec., 31 hrs. @ 12 sec. |
| Safety Standards: |
EN60601-1, (IEC 60601-1), EN60601-1-2, EN60601-1-1, UL 544, CSA Std. C22.2 No. 125-1984 |
| Regulatory Compliance / Safety Approvals: |
510(k): K970209 |
| ISO 9001 / EN46001 |
| Catalog Numbers: |
FGI-9004-050 Capnocheck® Sleep Capnograph |
| FGI-9004-051 Capnocheck® Sleep Capnograph/Oximeter |
| Go to Capnograph Specifications |
| SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE |