| FGI Oximeter Specifications (Nominal) |
| |
Autocorr® Digital Pulse Oximeter |
|
| FGI-3304 |
| Displays, Indicators & Keys: |
| Display |
LED numeric display |
| Resolution |
|
| Waveform |
| Graphic Trends |
| SpO2 |
3-digit LED numeric display |
| 0.43 inch (10.9 mm) high |
| Pulse Rate |
3-digit LED numeric display |
| 0.43 inch (10.9 mm) high |
| Pulse Strength |
Logarithmically scaled 10-segment LED bar graph |
| Low Battery |
LED flashes when approximately 30 minutes of battery life remains |
| Alarm Silence |
LED flashes or lights to remind the user that audio alarms are disabled |
| Sensor |
LED flashes to alert the operator to check the SpO2 probe placement |
| Lost Pulse Alarm |
Flashing % SpO2 dashes indicate the sensor no longer detects a pulse while a finger is inserted in it (a pulse must previously be detected in order to get this alarm) |
| Artifact |
Light indicates the presence of excess motion or other artifact |
| Power |
Green color indicates the monitor is connected to AC power |
| SpO2: |
| Range |
0-100% |
| Accuracy |
±2% at 70-100% |
| ±3% at 50-69% |
| Averaging |
4, 8, or 16 pulse beat average |
| Alarm Range |
High-50%-100% & OFF (1% steps) |
| Low-OFF & 50%-99% (1% steps) |
| Pulse Rate: |
| Range |
30-254 bpm |
| Accuracy |
±2 bpm or ±2% (whichever is greater) at 30-254 bpm |
| Averaging |
8 or 16 second average |
| Alarm Range |
High: 5-250 & OFF (1 bpm steps) |
| Low: 5-250 & OFF (1 bpm steps) |
| Power Requirements: |
105-125 V~, 60Hz (Wall Mount) |
| 100 V~, 50Hz (optional) |
| 230 V~, 50/60Hz (Table Top) |
| Battery Life: |
Standard Lead-Acid battery: |
| -Internally rechargeable; not user replaceable |
| -Fully charges in about 6 hours |
| -Approximately 4.5 hours of continuous use |
| Dimensions: |
3.24" H x 8.5" W x 5.5" D |
| 216 mm x 82 mm x 140 mm |
| Weight: |
30 oz. (850 grams) |
| Environment:; |
| Operating Temperature |
32°F to 104°F (0°C to 40°C) |
| Storage Temperature |
-40°F to 167°F (-40°C to 75°C) |
| Relative Humidity |
10-95% (storage) |
| (non condensing) |
15-95% (operating) |
| Auxiliary Serial Printer/Computer Output: |
SpO2 and pulse rate can be printed every five (5) seconds (data log) |
| Data saved every four (4) to thirty (30) seconds can be printed (trend) |
Analog Output (optional cable)
Digital Output (optional cable)
Nurse Call (optional cable) |
| Safety Standards: |
EN60601-1, EN60601-1-1, UL544, CSA std C22.2 No. 125-1984 |
| Catalog Numbers: |
FGI-3304 |
| SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE |