| FGI-6004 Specifications |
| Displays, Indicators & Keys: |
|
SpO2 (Optional): |
| SpO2 (Optional) |
3-digit LED display, 0.40 inches (10.16 mm) high |
Range |
0 to 100 SpO2 (functional) |
| Pulse Rate |
3-digit LED display, 0.30 inches (7.62 mm) high |
Accuracy |
±2 counts at 70 to 100%
<70% unspecified |
| Pulse Strength |
Logarithmically scaled 10-segment LED bar graph |
Averaging |
8 beats |
| NIBP |
3-digit LED display
|
Pulse Tone |
Volume adjustable or OFF
(pitch corresponds to SpO2 value) |
| Systolic & Diastolic Pressures - |
0.40 inches (10.16 mm) high |
Alarm Defaut |
Adult: |
High: 150 bpm
Low: 45 bpm |
| Mean Arterial Pressure - |
0.30 inches (7.62 mm) high |
Neonate: |
High: 150 bpm
Low: 50 bpm |
| Temperature |
3-digit LED display, 0.40 inches (10.16 mm) high |
Temperature (Optional): |
| NIBP: |
Range |
84°F to 108°F (28.9°C to 42.2°C) |
| Range |
Adult: |
20 to 250 mmHg |
Accuracy |
±2°F, ASTM E1112-86 (±0.1°C)
±3 counts at 50 to 69% |
| Neonate: |
20 to 135 mmHg |
Calibration |
Factory calibrated |
| NIBP Accuracy |
Meets SP10-1992 AAMI standard for NIPB accuracy |
Dimensions: |
| Cuff Pressure Accuracy |
±3 mmHg or 2% (whichever is greater) |
Base Unit |
3.24” H x 8.5” W x 5.5” D |
| Determination Time |
30 to 50 seconds typical (max. 120 seconds in adult mode, 90 seconds in neonate mode) |
8.2 cm H x 21.6 cm W x 14.0 cm D |
| Interval Times |
1, 2, 2.5, 5, 10, 15, 20, 30, 45, 60, 90 mins |
With Temp |
3.24” H x 10” W x 5.5” D |
| Calibration |
Factory calibrated |
8.2 cm H x 25.4 cm W x 14.0 cm D |
| Inflation Pressure |
Adult: |
50 to 300 mmHg |
Weight: |
| Neonate: |
50 to 150 mmHg |
Base Unit |
3.5 pounds (1.6 kg) |
| Default Inflation Pressure |
Adult: |
175 mmHg |
With Temp |
4.2 pounds (1.9 kg) |
| Neonate: |
100 mmHg |
AC Charger Power Requirements |
| Alarm Range |
105 to 125 VAC, 60 Hz |
| Adult |
High: |
15 to 285 mmHg & OFF (1 mmHg steps) |
90 to 110 VAC, 50 Hz (optional); 230 VAC, 50/60 Hz (optional) |
| Low: |
10 to 280 mmHg & OFF (1 mmHg steps)
|
Output of 24VDC @ 1.5A with 4kV isolation |
| Neonate |
High: |
15 to 135 mmHg & OFF (1 mmHg steps) |
Battery Life |
| Low: |
10 to 130 mmHg & OFF (1 mmHg steps)
|
Nickel Cadmium battery, 6 VDC, internal rechargeable, fully charged continuous-life 6 hours approximate, charge time 4 hours |
| Alarm Default |
Printer Output: |
| Adult |
Systolic: |
High -200 mmHg |
RS232C |
| Low -90 mmHg |
Data Format: 19200 baud, 1 start bit, 8 data bits, 1 stop bit, no parity |
| Diastolic: |
High -105 mmHg |
Integrated Printer (Optional): |
| Low -40 mmHg |
Text only, no graphics. Patient data log or trend tables. |
| Mean Arterial: |
High -140 mmHg |
Environmental: |
| Low -50 mmHg |
Operationg Temperature |
| Pulse Rate: |
Base Unit |
32°F to 131°F (0°C to 55°C) |
| Range |
30 to 180 bpm (NIBP) (adult) |
With Temp |
60.8°F to 104°F (16°C to 40°C) |
| 40 to 180 bpm (NIBP) (neonate) |
Storage Temperature |
-40°F to 158°F (-40°C to 70°C) |
| 30 to 254 bpm (SpO2) |
Relative Humidity |
10 to 95% (storage) |
| Accuracy |
±2% or 2 bpm (whichever is greater) |
(non-condensing) |
15 to 95% (operating) |
Alarm Range
Adult & Neonate |
High: 30 to 250 bpm & OFF (1bpm steps) |
Standards: |
| Low: 30 to 250 bpm & OFF (1bpm steps) |
EN60601-1 (IEC60601-1), EN60601-1-1 (IEC60601-1-1) UL 544, CSA Std. C22.2 No. 125-1984, EN60601-1-2 (IEC60601-1-2) |
| Alarm Default |
Adult: |
High: 150 bpm |
| Low: 45 bpm |
Regulatory Compliance: |
| Neonate: |
High: 150 bpm |
510(k): K983796, K984618, K970801 ISO 9001 / EN46001 |
| Low: 50 bpm |
|
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE
|