| C-type Super Open MRI System with Permanent Magnet |
 |
FGI-NAM-P023A
|
 |
|
| Specifications |
| Magnet: |
| Magnet type |
c-type 2-column super open, eddy current free, Permanent Magnet |
| Magnet strength |
2300 ± 50GS |
| Homogeneity |
20ppm in 30cm DSV |
| B0 Direction |
Vertical |
| Dimensions |
195x163x154 (LxWxH) |
| Weight |
13000kg |
| RF Transmitter System: |
| Operating Frequency |
9.8MHz ± 0.5MHz |
| RF Power |
5kw |
| Transmit Coil |
Plate type |
| Gradient System: |
| Maximum Gradient Strength |
12mT/m |
| Rise Time |
250 micro seconds |
| Maximum Slew Rate |
47mT/m/ms |
| Gradient Coil Type: |
| The plate type gradient coil based on the dactylogram design and computer control, hydro manufacture assure high performance and lowest noise during the scan, providing patients with comfortable examination conditions. |
| Scan Specifications: |
| TE |
Minimum 6ms Maximum 300ms |
| TR |
Minimum 20ms |
| Slice Thickness |
Minimum 3mm(2D), 1mm(3D) |
| Slice Numbers |
Maximum 64(2D) Maximum 128(3D) |
| FOV |
Minimum 50mm Maximum 400mm |
| Scan Matrix |
(64 x 1024) x (32 x 512) |
| Reconstruction Matrix |
(256 x 256) x (512 x 512) x (1024 x 1024) |
| Imaging |
- Multi-angle/multi-slice scanning: Multi-group with multi-angle/multi-slices in one scan protocol make the imaging of neck/spine/L-spine and c-spine much easier and more flexible
- Fractional Echo: Fractional echo technique shortens the echo time, increases the T1 contrast, and reduces the effect of the T2 at the same time, increasing the slice number in the given TR. Half scanning/reconstruction reduces the scan time
- Flow Compensation Technique: FGI-NAM-P023A flow compensation technique can remove the flow artifacts, making the image of the spine more clear
- Anti-aliasing Techniques: Frequency/phase oversample can remove the aliasing artifacts, obtaining excellent quality small FOV images
- Multi-region Pre-saturation Technique:
- Drawing multi-regions are flexible and can be saturated on the pilot image to remove or minimize the artifacts from breathing or abnormal signal of the ROI
- Optimized bandwidth according to the TE, samples points in sequence automatically, ensure to get the highest SNR images
- Optimized data acquisition including frequency under/over sampling and echo asymmetry techniques
- Gated-scanning technique
- ECG-gated scanning
- Provides many combinations of the phase encoding sequence, suitable to different scanning situations such as the critically ill or patients who cannot remain still for a long scanning period
|
| Computer System |
- Two Computer Systems Based on TCP/IP Connection
- Provide all kinds of functions such as image analysis, process, print and remote-interoperation during the scanning.
- Point and click operation of the mouse can realize most functions of the system.
- Pre-Registration Of The Patients
- User may construct the list of patients or re-order the queue to save scan time.
- Patient information management, authorized scan and safety management.
- Main Computer System:
- Pentium III, 700MHZ, windows NT workstation
- 128MB memory (can be extended to 1024MB)
- >2020GB
- Mo Storage:
- 640MB MO storage, user can view images, analyze and make statistic in the MO itself or transfer the data between the hard disk and MO easily.
- Monitor:
- 21" high resolution color display
- Reconstruction Time:
- Network:
- Ethernet network based on TCP/IP protocols, windows NT system.
- Two computer systems are connected by Ethernet to perform remote access to the system.
- Spectrometer:
- Based on windows NT system, industrial control PC, realizing scanning and reconstruction simultaneously.
- Scan can be interrupted at any time.
- Laser Printer Interface:
- Digital interface, compatible to DICOM3.0; Can be connected to most well-known laser printers or to the devices in conformity to standard DICOM3.0
- Patient Table:
- Progress distance 160cm
- Maximum patient load:180kg
- Moving styles: two dimension, manual-driven
- Patient Table Location:
- Cross laser line for location
- Patient Monitor:
- Intercom, Background music
|
| Multiple Receiver Coils |
- FGI-NAM-P023A features a rich lineup of coils. Each of them provides patient-friendliness, easy-operation and advanced technology for high image quality.
- Head coil and body coil are both QD two-channels. SNR can be improved by about 40% higher than conventional solenoid coil.
- Flexible body boil can fit different shaped patients as well as improve fill factor to get better image.
- Flat coil for spine imaging dramatically increases the clarity of image in ROI.
- FGI-NAM-P0234A offers a variety of coils with coil designs to fit anatomical regions.
- Coils Offered:
- Neck Coil
- Flat Spine Coil
- QD Head Coil
- Flexible Coil
- Breast Coil
- QD Chest Coil
- Joint Coils for Knee, Extremity, etc.
|
| Friendly User Interface, Easy to Operate |
- User interface: well designed user interface makes it easy to be trained
- Voice: high quality intercom and background music, makes patients feel at ease
- Graphical Functional Key Design
- Combination of floating cards and fixed function area makes the operation easy, flexible and fast.
- Full Set Of Pulse Sequence Library, Including Most Clinical Sequences
- User can point the standard scan protocols to start a scan immediately, or modify the parameters.
- Multi-image Processing Functions
- Magnify, zoom, rotation, combination, transpose, cut/paste, dynamic cine display, mark, continuous windowing, drawing point/line/area, measure the length/area, roi statistic etc.
- Low Runing Cost, High And Quick Return On Investment
- Premium design magnet with powerful permanent material, minimizing space required, no need to re-fill the liquid helium and cooling water, lowest operating and maintenance cost, ensure quick and high return on investment.
- Continuous Upgrade Possibility
- As part of our continous R&D you will have access to the latest upgrades
|
| Site Requirements |
| Scan Room: |
| Temperature |
21±2°C |
| Relative Humidity |
40%~80% |
| Minimum Space |
4m(W) x 6.5m(L) x 2.5m(H) |
| RF Shielding Facility |
Attenuation > -80dB in 1MHz-100MHz |
| Machine Room: |
| Temperature |
21±5°C |
| Relative Humidity |
40%~80% |
| Minimum Space |
3m(W) x 2.5m(L) x 2.5m(H) |
| Demand for Lithometer |
No Dust |
| Operation Room: |
| Temperature |
21±5°C |
| Relative Humidity |
40%~80% |
| Minimum Space |
3m(W) x 2m(L) x 2.5m(H) |
| Power Requirements: |
| Type |
3-phase, 4-wire |
| Capacity |
10KVA |
| Frequency |
50Hz+1% |
| Voltage |
380V/220V±10% |
| Grounding Facility |
Grounding resistance should be less than 1Ω and exclusive for MRI system. |
|
|
|