| FGI Hand-Held Oximeter Specifications |
| Capnocheck® Hand-Held Capnograph/Oximeter |
 |
| FGI-8400 |
| Displays, Indicators & Keys: |
| SpO2 |
LCD, with EL back light |
| Graphical display for ETCO2, respiration rate, %SpO2, or pulse rate; user selectable scale |
| Pulse Rate |
| CO2, waveform or oximetry plethysmogram, user selectable |
| Pulse Strength |
8-segment bargraph indicates logarithmic strength of patient's pulse. Not proportional to pulse volume. |
| Alarm Silence |
High and medium/low priority alarm, Alarms silenced, and AC power |
| SpO2: |
| Range |
0 to 100% SpO2 (functional) |
| Accuracy |
±2% at 70 to100% |
| less than 70% unspecified |
| Averaging |
8 beats |
| Pulse Rate: |
| Range |
30 to 254 bpm |
| Accuracy |
±2 bpm or ±2% whichever is greater |
| Averaging |
8 beats |
| Power Requirements: |
120V~, 60Hz
|
| 230V~, 50/60Hz |
| 100V~, 60 Hz (Optional) |
| Battery Life: |
MoliEnergy Battery Charger |
| Ll+ (lithium-Ion), 7.4 VDC Battery |
| Replaceable, internal rechargeable. |
| Fully-charged continuous life of approximately 6 hours. Maximum full-capacity charging is 2.5 hours. |
| Dimensions: |
5.0"H x 4.38"W x 2.9"D |
| 122mm H x 111mm W x 73.7mm D |
| Weight: |
22 ounces (0.63 Kg |
| Environment:; |
| Operating Temperature |
32°F to 122°F (0°C to 50°C) |
| Storage Temperature |
-40°F to 167°F (-40° to 75°C) |
| Relative Humidity |
10 to 95% (storage) |
| (non condensing) |
15 to 95% (operating) |
| Auxiliary Serial Printer/Computer Output: |
Infrared Port compatible with HP82240B Printer or similar device. |
| Non-standard data format ~ 500 baud |
| Text only, no graphics. User-selectable patient data log or trend tables. |
| Mechanical Durability Test Results: |
| Standards: |
EN60601-1 (IEC 60601-1) |
| UL544, CSA Std. C22.2 No. 125-1984 |
| EN60601-1-2 (IEC 60601-1-2) |
| Regulatory Compliance / Safety Approvals: |
510(k): K991086 |
| ISO 9001 / EN 46001 |
| Catalog Numbers: |
FGI-8400 Capnograph/Oximeter |
| FGI-8401 Capnograph |
| Go to capnograph specifications |
| SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE |