| FGI Hand-Held Oximeter Specifications |
| Digit Finger Oximeter |
|
| FGI-3420 |
| Displays, Indicators & Keys: |
| SpO2 |
LED numeric display |
| 0.32 inches (8.1 mm) high |
| Pulse Rate |
LED numeric display |
| 0.32 inches (8.1 mm) high |
| Pulse Strength |
LED bar graph display 0.32 inches (8.1 mm) high |
| Keys |
On key |
| SpO2: |
| Range |
0 to 99% |
| Accuracy |
±2% at 70 to 99% |
| less than 70% unspecified |
| Averaging |
8 pulse beat average |
| Pulse Rate: |
| Range |
30 to 254 bpm |
| Accuracy |
±2% or 2 bpm (whichever is greater) |
| Averaging |
8 second average |
| Power Requirements: |
(2) "AAA" Cell Alkaline Batteries (IEC Type LR03) |
| Battery Life: |
16 hours continuous use |
| 1400 spot checks, 1 min. on, 2 min. off |
| Dimensions: |
1.7"W x 2.25"H x 1.5"D |
| (43.2mm x 57.2mm x 38.1mm) |
| Weight: |
3 oz. (85 grams) with batteries |
| Environment:; |
| Operating Temperature |
32°F to 131°F (0°C to 55°C) |
| Storage Temperature |
-40°F to 167°F (-40° to 75°C) |
| Relative Humidity |
10 to 95% (storage) |
| (non condensing) |
15 to 95% (operating) |
| Standards: |
EN60601-1 (IEC60601-1) |
| UL2601-1 and CAN/CSA 601.1 |
| CAN/CSA NRTL/C |
| Regulatory Compliance / Safety Approvals: |
510(k); K013171 |
| ISO 9001 / EN 46001 |
| Catalog Numbers: |
FGI-3420Y Digit® Finger Oximeter, Yellow |
| FGI-3420G Digit® Finger Oximeter, Gray |
| SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE |