| FGI Hand-Held Oximeter Specifications |
| FingerPrint® Hand-Held Pulse Oximeter |
|
| FGI-3401 |
| Displays, Indicators & Keys: |
| SpO2 |
LED numeric display |
| 0.300 inch (7.62mm) high |
| Pulse Rate |
LED numeric display |
| 0.300 inch (7.62mm) high |
| Pulse Strength |
Logarithmically scaled, 8-segment bargraph |
| Keys |
On/Off, Print and Speaker On/Off |
| SpO2: |
| Range |
0-99% |
| Accuracy |
±2% at 70-99% |
| <70% unspecified |
| Averaging |
Averaging 8 beat average |
| Pulse Rate: |
| Range |
30-254 bpm |
| Accuracy |
±2 bpm or ±2%, whichever is greater |
| Averaging |
8 second average |
| Power Requirements: |
(4) "AA" Alkaline or NiCAD Batteries |
| Battery Life: |
24 hours without printing, 14 hours of continuous |
| printing, 2000 spot checks |
| Dimensions: |
6.6" H x 2.75" W x 1.43" D |
| 16.7 cm x 7.0 cm x 3.6 cm |
| Weight: |
13 oz. (369 grams) with batteries |
| Environment:; |
| Operating Temperature |
32°F to 131°F (0°C to 55°C) |
| Storage Temperature |
-40°F to 167°F (-40°C to 75°C) |
| Relative Humidity |
10-95% (storage) |
| (non condensing) |
15-95% (operating) |
| Memory / Trend Storage |
SpO2, Pulse rate and Relative Time |
| 14 hours of storage at 30-second intervals |
| Auxiliary Serial Printer/Computer Output: |
RS232C |
| Integrated Printer: |
| 1-inch wide thermal printer paper |
| 15 characters per line |
| Mechanical Durability Test Results: |
| Vibration Testing |
In compliance with IEC 68-2-6, IEC 68-2-34 |
| MIL-STD-810E, Method 514.4 |
| Shock Testing |
In compliance with IEC 68-2-27 |
| Drop Testing |
IEC 60601-1, Section 21.5 |
| Standards: |
EN60601-1, (IEC60601-1), UL544, |
| CSA Std. C22.2 No.125-1984 |
| EN 60601-1-2 |
| Regulatory Compliance / Safety Approvals: |
510(k): K980714 |
| ISO 9001 / EN46001 |
| Catalog Numbers: |
FGI-3401 |
| SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE |