Fax Transmission Function
Direct faxing (from the ECG) prints much clear ECG reports at the receiving end. If an external modem is connected with CardioCare2000, user can transmit ECG signal to remote area easily and receive any message with ECG machine.
Enhanced interpretation
With professional interpretation result, ECG printout offers detailed interpretation information in beat report format, which is especially helpful to doctors making critical decisions. |
| FGI-CardioCare2000 |
|
| S P E C I F I C A T I O N S |
| ECG Leads |
12 leads simultaneous 10-second resting ECG |
| Dimension |
296(W) x 305.5(H) x 92.5(D)mm, Approx. 2.98kg |
| Recording Channel |
3,6,12 channel and 1 channel(60 sec) |
| Sensitivity |
5, 10, 20, auto(I~aVF: 10, V1~V6: 5) mm/mV |
| Printing Speed |
12.5, 25, 50 mm/S |
| Filters |
AC(50/60Hz, -20dB or better)
Muscle(25 - 35HZ, -3dB or better)
Base line drift (0.1Hz, -3dB or better)
Low pass filter (off, 40Hz, 100Hz, 150Hz) |
Display
LCD Display
LED/Monitor Display |
2 x 16 character LCD Display
Sensitivity, speed, filter status, HR, printing form, rythm lead
signal quality, power source, battery |
| Manual Interface |
Membrane keyboard, Alphanumeric and symbol available |
Patient Data
Basic Measurement |
ID, name, age, sex, height, weight
Heart rate, PR, QRS, QT/QTc, P-R-T axis |
| Recorder |
Thermal print head, Thermal roll paper Report papers:
width: A4: 210mm or 8.5"
length: A4: 300mm or 11"
Effective writing width: 204mm Resolution:
Vertical: 8 dots/mm
Horizontal: 16 dots/mm |
| Electrical |
Internal noise: 20µV(p-p)max
Input circuit: Floating input
Isolated and defilibration protected
Input impedance: ≥10MΩ
Input range: ≥±5mV
Common Mode Rejection: ≥ +100dΒ
DC offset voltage: ≥±300mV
Time constant: 3.2sec
Patient leakage current: <10μA
Frequency response 0.05~150Hz |
| Signal Quality Control |
Disconnected lead detection |
| Power |
Power supply: AC or built-in Battery(option)
Power Requiement: 100-240 Vac, 50/60Hz, 1.0-0.5 A
Power consumption: 60W max |
| Battery Capacity |
1 hour of normal use (approx. 100 automatic ECG printouts) |
| Communication |
PC connection with RS-232 interface and LAN |
Safety
Conformity |
Class I, Type BF
CE, CSA |
| Environment |
Operating Humidity: 30 ~ 85% Operating Temperature: 10° ~ 40°C Atmospheric Pressure: 70 ~ 106 KPa |
| Standard Accessory |
Power Cable 1ea, Patient Cagle 1ea, Electrode 10ea, Chart Paper 1ea, Operation Manual 1ea |
| Options |
Rechargeable Battery(NI-MH), Cart, Hanger, Bag, PC s/w, Limb electrode, Chest electrode, RS-232C null cable, LAN cable(Ethernet RJ-45) |
| SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE |
|