| FGI Capnometer / Capnograph Specifications (Nominal) |
| |
Capnocheck® Sleep Capnograph/Oximeter |
Capnocheck® II Capnograph |
Capnocheck® Plus Capnograph |
|
 |
|
| FGI-9004-050 |
FGI-8401 |
FGI-9004 |
| Displays, Indicators & Keys: |
| Display |
Vacuum Fluorescent |
LCD, with EL back light |
Vacuum Fluorescent |
| Indicators |
Alarm, Alert, Charge, Silenced |
High and medium/low priority alarm, Alarms silenced, and AC power |
Alarm, Alert, Charge, Silenced |
| Displayed Waveforms |
Selectable CO2 or CO2/Plethysmogram waveform |
CO2, waveform or oximetry plethysmogram, user selectable |
Selectable CO2 or CO2/Plethysmogram waveform |
| Graphic Trends |
Graphical display, factory selectable sample rate, user selectable trend up to 16 hr. display @ 12 sec. sample rate up to 10 hr. display @ 4 sec. sample rate |
Graphical display for ETCO2, respiration rate, %SpO2, or pulse rate; user selectable scale |
Graphical display, factory selectable sample rate, user selectable trend 16 hr. display @ 12 sec. sample rate, 10 hr. display @ 4 sec. sample rate |
| Alarms: |
| Physiological |
Adjustable volume user-selectable high and low limits for all displayed parameters
Screen indicators for each parameter |
User-selectable high and low limits, adjustable volume, screen indicators |
Adjustable volume User-selectable high and low limits for all displayed parameters
Screen indicators for each parameter |
| System |
Adjustable volume |
Adjustable volume |
Adjustable volume |
| Silence |
Temporary 2 min. or OFF LED indicator |
Temporary 2 min. or infinite, LED indicator |
Temporary 2 min. or OFF LED indicator |
| Capnograph: |
| Flow Rate |
150 ml/min, ± 20 |
120±20 ml/min |
120 ml/min, +/-20 |
| Measurement |
Non-Dispersive IR absorption |
Non-dispersive IR absorption |
Non-Dispersive IR absorption |
| MeasurementRange |
0-10% CO2 |
0 to 10% CO2 STPD (standard temperature and pressure dry) |
0-10% CO2 |
| Units |
User-selectable: mmHg, Vol. %, kPa |
|
User-selectable: mmHg, Vol. %, kPa |
| Display Range |
0-100 mmHg or 0-10.0% CO2 or 0-13.3 kPa |
0 to 100 mmHg; 0-13 kPa; 0-10% CO2 |
0-100 mmHg or 0-10.0% CO2 or 0-13.3 kPa |
| Accuracy |
±2 mmHg; ±0.3%; ±0.3 kPa, or 4% of reading, whichever is greater (0-10%CO2) |
±2mmHg or 4% of reading, whichever is greater |
±2 mmHg; ±0.3%; ±0.3 kPa, or 4% of reading, whichever is greater (0-10% CO2) |
| Resolution |
1 mmHg; 0.1%; 0.1 kPa |
|
1 mmHg; 0.1%; 0.1 kPa |
| Calibration |
2 point Cal |
|
2 point Cal |
| N2O Compensation |
User-selectable 40% (default=OFF) |
selectable 40% (default = OFF) |
User-selectable 40% (default = OFF) |
| Response Time |
375 msec to 90% of value |
2.46s to 90% with 8 ft. patient line |
375 msec to 90% of value |
| Averaging |
4-breath average, or breath-to-breath selectable |
4-breath average |
4-breath average or breath by breath, user selectable |
| Respiration Rate: |
| Range |
0-150 bpm |
0 to 150 breaths/min |
0-150 bpm |
| Averaging |
4 breath average |
4 breath average |
4-breath average or breath by breath, user selectable |
| Accuracy |
± 1bpm |
± 1bpm |
± 1bpm |
| SpO2: |
| Range |
0-100% |
0 to 100% SpO2 (functional) |
0-100% |
| Accuracy |
±2% @ 70-100% |
±2% at 70 to100% |
±2% at 70-100% |
| <70% unspecified |
less than 70% unspecified |
±3% at 50-69% |
| Averaging |
2, 4, 8, or 16-beat averaging (default: 2-beat) |
8 beats |
4, 8 or 16 beat, user-selectable |
| Pulse Rate: |
| Range |
30-254 bpm |
30 to 254 bpm |
30-254 bpm |
| Accuracy |
±2 bpm or ±2%, whichever is greater |
±2 bpm or ±2% whichever is greater |
±2 bpm or ±2%, whichever is greater |
| Averaging |
8 second average |
8 beats |
8 or 16 seconds, user-selectable |
| Pulse Tone |
Pitch corresponds SpO2 value
Volume adjustable or OFF |
|
Pitch corresponds to SpO2 value
Volume adjustable or OFF |
| Power Requirements: |
| AC Charger Input |
115V~, 60 Hz |
120V~, 60Hz; 230V~, 50/60Hz |
105-125 V~, 60Hz (Wall Mount) |
| (Optional) 100V~, 50 Hz |
230V~, 50/60Hz |
100 V~, 50Hz (optional) |
| (Optional) 230V~, 50/60 Hz |
100V~, 60 Hz (Optional) |
230 V~, 50/60Hz (Table Top) |
| Battery: |
Nickel Cadmium battery, 6 VDC, internal rechargeable, not user replaceable, fully charged continuous- use life 2 hours approximate, maximum full charge time 4 hours |
MoliEnergy Battery Charger
Ll+ (lithium-Ion), 7.4 VDC Battery Replaceable, internal rechargeable. Fully-charged continuous life of approximately 6 hours. Maximum full-capacity charging is 2.5 hours. |
Nickel Cadmium battery, 6 VDC, internal rechargeable, not user replaceable, fully charged continuous - use life 2 hours approximate, maximum full charge time 4 hours |
| Dimensions: |
3.5" H x 10.0" W x 5.5" D |
5.0"H x 4.38"W x 2.9"D |
3.5" H x 10.0" W x 5.5" D |
| 88.9 mm x 254 mm x 139.7 mm |
122mm H x 111mm W x 73.7mm D |
88.9 mm x 254 mm x 139.7 mm |
| Weight: |
4.8 pounds (2.2 kg) |
22 ounces (0.63 Kg |
5 lbs. (2.27 kg) |
| Environment:; |
| Operating Temperature |
32°F - 122°F (0°C - 50°C) |
32°F to 122°F (0°C to 50°C) |
32°F to 131°F (0°C to 55°C) |
| Storage Temperature |
-40°F-158°F (-40°C-70°C) |
-40°F to 167°F (-40° to 75°C) |
-40°F to 167°F (-40°C to 75°C) |
| Relative Humidity |
10 - 95% (storage) |
10 to 95% (storage) |
10-95% (storage) |
| (non condensing) |
15 - 95% (operating) |
15 to 95% (operating) |
15-95% (operating) |
| Analog Output: |
Any four of: CO2 waveform, Plethysmogram, End Tidal CO2, Resiration Rate, Inspired CO2, % SpO2, Pulse Rate, 0V or 1V Cal Signal
Range: 0-1 VDC full scale |
Any three of: CO2 waveform, Plethysmogram, End Tidal CO2 Respiration Rate, Inspired CO2, %SpO2, Pulse Rate Fractional Inspired Oxygen, 0V or 1V Cal signal
Range 0-1 VDC full scale |
|
| Serial Output: |
RS232C: tabular trends, patient data
For Sleep Study: 10 hrs. @ 4 sec., 31 hrs. @ 12 sec. |
Infrared Port compatible with HP82240B Printer or similar device.
Data Format: Non-standard data format ~ 500 baud
Options: Text only, no graphics. User-selectable patient data log or trend tables. |
RS232C: tabular trends, patient data
For Sleep Study: 10 hrs. @ 4 sec., 31 hrs. @ 12 sec. |
| Safety Standards: |
EN60601-1, (IEC 60601-1), EN60601-1-2, EN60601-1-1, UL 544, CSA Std. C22.2 No. 125-1984 |
EN60601-1 (IEC 60601-1) UL544, CSA Std. C22.2 No. 125-1984 EN60601-1-2 (IEC 60601-1-2) |
EN60601-1, (IEC60601-1), UL544, CSA Std. C22.2 No.125-1984, EN 60601-1-2 |
| Regulatory Compliance / Safety Approvals: |
510(k): K970209 |
510(k): K991086 |
|
| ISO 9001 / EN46001 |
ISO 9001 / EN 46001 |
|
| Catalog Numbers: |
FGI-9004-050 Capnocheck® Sleep Capnograph |
FGI-8400 Capnograph/Oximeter |
FGI-9004 Capnocheck® Plus Capnograph |
| FGI-9004-051 Capnocheck® Sleep Capnograph/Oximeter |
FGI-8401 Capnograph |
|
| |
Go to Oximeter Specifications |
Go to Oximeter Specifications |
|
| * * * Highlighted indicates specifications of optional items * * * |
| SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE |