First Gulf International, Inc.
First Gulf International, Inc.
Site Map
Complete Line of Medical Equipment & Hospital Package Deals
spacer1
spacer2

FGI-9004 Capnocheck® Plus Capnograph

FOR PRICING AND DELIVERY, CALL US AT 713-961-7793 OR EMAIL US



FGI Capnometer / Capnograph Specifications (Nominal)
Capnocheck® Plus Capnograph
FGI-9004 Capnocheck Plus Capnograph 
FGI-9004
Displays, Indicators & Keys:
Display Vacuum Fluorescent
Indicators Alarm, Alert, Charge, Silenced
Displayed Waveforms Selectable CO2 or CO2/Plethysmogram waveform
Graphic Trends Graphical display, factory selectable sample rate, user selectable trend 16 hr. display @ 12 sec. sample rate, 10 hr. display @ 4 sec. sample rate
Alarms:
Physiological Adjustable volume User-selectable high and low limits for all displayed parameters
Screen indicators for each parameter
System Adjustable volume
Silence Temporary 2 min. or OFF LED indicator
Capnograph:
Flow Rate 120 ml/min, +/-20
Measurement Non-Dispersive IR absorption
Measurement–Range 0-10% CO2
Units User-selectable: mmHg, Vol. %, kPa
Display Range 0-100 mmHg or 0-10.0% CO2 or 0-13.3 kPa
Accuracy ±2 mmHg; ±0.3%; ±0.3 kPa, or 4% of reading, whichever is greater (0-10% CO2)
Resolution 1 mmHg; 0.1%; 0.1 kPa
Calibration 2 point Cal
N2O Compensation User-selectable 40% (default = OFF)
Response Time 375 msec to 90% of value
Averaging 4-breath average or breath by breath, user selectable
Respiration Rate:
Range 0-150 bpm
Averaging 4-breath average or breath by breath, user selectable
Accuracy ± 1bpm
SpO2:
Range 0-100%
Accuracy ±2% at 70-100%
±3% at 50-69%
Averaging 4, 8 or 16 beat, user-selectable
Pulse Rate:
Range 30-254 bpm
Accuracy ±2 bpm or ±2%, whichever is greater
Averaging 8 or 16 seconds, user-selectable
Pulse Tone Pitch corresponds to SpO2 value
Volume adjustable or OFF
Power Requirements:
AC Charger Input 105-125 V~, 60Hz (Wall Mount)
100 V~, 50Hz (optional)
230 V~, 50/60Hz (Table Top)
Battery: Nickel Cadmium battery, 6 VDC, internal rechargeable, not user replaceable, fully charged continuous - use life 2 hours approximate, maximum full charge time 4 hours
Dimensions: 3.5" H x 10.0" W x 5.5" D
88.9 mm x 254 mm x 139.7 mm
Weight: 5 lbs. (2.27 kg)
Environment:;
Operating Temperature 32°F to 131°F (0°C to 55°C)
Storage Temperature -40°F to 167°F (-40°C to 75°C)
Relative Humidity 10-95% (storage)
(non condensing) 15-95% (operating)
Serial Output: RS232C: tabular trends, patient data
For Sleep Study: 10 hrs. @ 4 sec., 31 hrs. @ 12 sec.
Safety Standards: EN60601-1, (IEC60601-1), UL544, CSA Std. C22.2 No.125-1984, EN 60601-1-2
Catalog Numbers: FGI-9004 Capnocheck® Plus Capnograph
* * * Highlighted indicates specifications of optional items * * *
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE

Powered by WebWorks USA - Website Hosting, Design and Promotion in Houston, Texas