| FGI Capnometer / Capnograph Specifications (Nominal) |
| Capnocheck® Plus Capnograph |
|
| FGI-9004 |
| Displays, Indicators & Keys: |
| Display |
Vacuum Fluorescent |
| Indicators |
Alarm, Alert, Charge, Silenced |
| Displayed Waveforms |
Selectable CO2 or CO2/Plethysmogram waveform |
| Graphic Trends |
Graphical display, factory selectable sample rate, user selectable trend 16 hr. display @ 12 sec. sample rate, 10 hr. display @ 4 sec. sample rate |
| Alarms: |
| Physiological |
Adjustable volume User-selectable high and low limits for all displayed parameters
Screen indicators for each parameter |
| System |
Adjustable volume |
| Silence |
Temporary 2 min. or OFF LED indicator |
| Capnograph: |
| Flow Rate |
120 ml/min, +/-20 |
| Measurement |
Non-Dispersive IR absorption |
| MeasurementRange |
0-10% CO2 |
| Units |
User-selectable: mmHg, Vol. %, kPa |
| Display Range |
0-100 mmHg or 0-10.0% CO2 or 0-13.3 kPa |
| Accuracy |
±2 mmHg; ±0.3%; ±0.3 kPa, or 4% of reading, whichever is greater (0-10% CO2) |
| Resolution |
1 mmHg; 0.1%; 0.1 kPa |
| Calibration |
2 point Cal |
| N2O Compensation |
User-selectable 40% (default = OFF) |
| Response Time |
375 msec to 90% of value |
| Averaging |
4-breath average or breath by breath, user selectable |
| Respiration Rate: |
| Range |
0-150 bpm |
| Averaging |
4-breath average or breath by breath, user selectable |
| Accuracy |
± 1bpm |
| SpO2: |
| Range |
0-100% |
| Accuracy |
±2% at 70-100% |
| ±3% at 50-69% |
| Averaging |
4, 8 or 16 beat, user-selectable |
| Pulse Rate: |
| Range |
30-254 bpm |
| Accuracy |
±2 bpm or ±2%, whichever is greater |
| Averaging |
8 or 16 seconds, user-selectable |
| Pulse Tone |
Pitch corresponds to SpO2 value
Volume adjustable or OFF |
| Power Requirements: |
| AC Charger Input |
105-125 V~, 60Hz (Wall Mount) |
| 100 V~, 50Hz (optional) |
| 230 V~, 50/60Hz (Table Top) |
| Battery: |
Nickel Cadmium battery, 6 VDC, internal rechargeable, not user replaceable, fully charged continuous - use life 2 hours approximate, maximum full charge time 4 hours |
| Dimensions: |
3.5" H x 10.0" W x 5.5" D |
| 88.9 mm x 254 mm x 139.7 mm |
| Weight: |
5 lbs. (2.27 kg) |
| Environment:; |
| Operating Temperature |
32°F to 131°F (0°C to 55°C) |
| Storage Temperature |
-40°F to 167°F (-40°C to 75°C) |
| Relative Humidity |
10-95% (storage) |
| (non condensing) |
15-95% (operating) |
| Serial Output: |
RS232C: tabular trends, patient data
For Sleep Study: 10 hrs. @ 4 sec., 31 hrs. @ 12 sec. |
| Safety Standards: |
EN60601-1, (IEC60601-1), UL544, CSA Std. C22.2 No.125-1984, EN 60601-1-2 |
| Catalog Numbers: |
FGI-9004 Capnocheck® Plus Capnograph |
| * * * Highlighted indicates specifications of optional items * * * |
| SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE |