| FGI Capnometer / Capnograph Specifications (Nominal) |
| Capnocheck® Sleep Capnograph/Oximeter |
|
| FGI-9004-050 |
| Displays, Indicators & Keys: |
| Display |
Vacuum Fluorescent |
| Indicators |
Alarm, Alert, Charge, Silenced |
| Displayed Waveforms |
Selectable CO2 or CO2/Plethysmogram waveform |
| Graphic Trends |
Graphical display, factory selectable sample rate, user selectable trend up to 16 hr. display @ 12 sec. sample rate up to 10 hr. display @ 4 sec. sample rate |
| Alarms: |
| Physiological |
Adjustable volume user-selectable high and low limits for all displayed parameters
Screen indicators for each parameter |
| System |
Adjustable volume |
| Silence |
Temporary 2 min. or OFF LED indicator |
| Capnograph: |
| Flow Rate |
150 ml/min, ± 20 |
| Measurement |
Non-Dispersive IR absorption |
| MeasurementRange |
0-10% CO2 |
| Units |
User-selectable: mmHg, Vol. %, kPa |
| Display Range |
0-100 mmHg or 0-10.0% CO2 or 0-13.3 kPa |
| Accuracy |
±2 mmHg; ±0.3%; ±0.3 kPa, or 4% of reading, whichever is greater (0-10%CO2) |
| Resolution |
1 mmHg; 0.1%; 0.1 kPa |
| Calibration |
2 point Cal |
| N2O Compensation |
User-selectable 40% (default=OFF) |
| Response Time |
375 msec to 90% of value |
| Averaging |
4-breath average, or breath-to-breath selectable |
| Respiration Rate: |
| Range |
0-150 bpm |
| Averaging |
4 breath average |
| Accuracy |
± 1bpm |
| SpO2: |
| Range |
0-100% |
| Accuracy |
±2% @ 70-100% |
| <70% unspecified |
| Averaging |
2, 4, 8, or 16-beat averaging (default: 2-beat) |
| Pulse Rate: |
| Range |
30-254 bpm |
| Accuracy |
±2 bpm or ±2%, whichever is greater |
| Averaging |
8 second average |
| Pulse Tone |
Pitch corresponds SpO2 value
Volume adjustable or OFF |
| Power Requirements: |
| AC Charger Input |
115V~, 60 Hz |
| (Optional) 100V~, 50 Hz |
| (Optional) 230V~, 50/60 Hz |
| Battery: |
Nickel Cadmium battery, 6 VDC, internal rechargeable, not user replaceable, fully charged continuous- use life 2 hours approximate, maximum full charge time 4 hours |
| Dimensions: |
3.5" H x 10.0" W x 5.5" D |
| 88.9 mm x 254 mm x 139.7 mm |
| Weight: |
4.8 pounds (2.2 kg) |
| Environment:; |
| Operating Temperature |
32°F - 122°F (0°C - 50°C) |
| Storage Temperature |
-40°F-158°F (-40°C-70°C) |
| Relative Humidity |
10 - 95% (storage) |
| (non condensing) |
15 - 95% (operating) |
| Analog Output: |
Any four of: CO2 waveform, Plethysmogram, End Tidal CO2, Resiration Rate, Inspired CO2, % SpO2, Pulse Rate, 0V or 1V Cal Signal
Range: 0-1 VDC full scale |
| Serial Output: |
RS232C: tabular trends, patient data
For Sleep Study: 10 hrs. @ 4 sec., 31 hrs. @ 12 sec. |
| Safety Standards: |
EN60601-1, (IEC 60601-1), EN60601-1-2, EN60601-1-1, UL 544, CSA Std. C22.2 No. 125-1984 |
| Regulatory Compliance / Safety Approvals: |
510(k): K970209 |
| ISO 9001 / EN46001 |
| Catalog Numbers: |
FGI-9004-050 Capnocheck® Sleep Capnograph |
| FGI-9004-051 Capnocheck® Sleep Capnograph/Oximeter |
| Go to Oximeter Specifications |
| * * * Highlighted indicates specifications of optional items * * * |
| SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE |